Fluke Metrology Range

Wide Temperature Range

9170    –45 °C  to  140 °C  /  –49 °F  to  284 °F
9171     –30 °C  to  155 °C  /  –22 °F  to  311 °F
9172       35 °C  to  425 °C  /    95 °F  to   797 °F
9173       50 °C  to  700 °C  /  122 °F  to  1292 °F

Fluke 9170

The Model 9170 achieves the lowest temperatures of the series, reaching –45 °C in normal room conditions. The 9170 is stable to ±0.005 °C over its full temperature range (up to 140 °C) and has 160 mm (6.3 in) of immersion depth. With axial uniformity of ±0.02 °C and radial uniformity of ±0.01 °C, this model delivers exceptional uncertainty budgets and is perfect for a variety of pharmaceutical and other applications.

Fluke 9171

If you need more depth, the Model Fluke 9171 provides 203 mm (8 in) of immersion over temperatures from –30 °C all the way to 155 °C with a full-range stability of ±0.005 °C. Just like the Fluke 9170, this dry-well has exceptional axial and radial uniformity. The display of the Fluke 9171 is calibrated to an accuracy of ±0.1 °C over its full range.

Fluke 9172

The Model Fluke 9172 provides temperatures from 35 °C to 425 °C with a calibrated display accurate to ±0.2 °C at 425 °C. In addition to exceptional accuracy, the 9172 is stable from ±0.005 °C to ±0.01 °C, depending on temperature. With 203 mm (8 in) of immersion, the Fluke 9172 significantly reduces stem conduction errors at high temperatures.

Fluke 9173

For work between 50 °C and 700 °C, the Model Fluke 9173 provides unmatched performance. The Fluke 9173 has a display accuracy of ±0.25 °C at 700 °C and an immersion depth of 203 mm (8 in). Stability and uniformity performance of this unit is enough to dramatically reduce uncertainty budgets for calibrations of thermometers at high temperatures.

Fc-917x_01b_c

Selection Guide

Fluke 9170 Fluke 9171 Fluke 9172 Fluke 9173
Range -45°C to 140°C +30°C to +155°C +35°C to +425°C +50°C to +700°C
Stability ±0.005°C ±0.005°C ±0.005°C to 100°C
±0.008°C to 425°C
±0.005°C to 700°C
±0.005°C to 100°C
±0.01°C to 425°C
±0.03°C to 700°C
Reference Probe YES YES YES YES
Accuracy with Built-In Reference Probe ±0.007°C at -100°C
±0.006°C at 0°C
±0.011°C at 155°C
±0.013°C at 225°C
±0.019°C at 425°C
±0.027°C at 661°C
±0.007°C at -100°C
±0.006°C at 0°C
±0.011°C at 155°C
±0.013°C at 225°C
±0.019°C at 425°C
±0.027°C at 661°C
±0.007°C at -100°C
±0.006°C at 0°C
±0.011°C at 155°C
±0.013°C at 225°C
±0.019°C at 425°C
±0.027°C at 661°C
±0.007°C at -100°C
±0.006°C at 0°C
±0.011°C at 155°C
±0.013°C at 225°C
±0.019°C at 425°C
±0.027°C at 661°C
Accuracy without External Reference Probe ±0.1°C ±0.1°C ±0.1°C to 100°C
±0.15°C to 225°C
±0.2°C to 425°C
±0.2°C to 425°C
±0.25°C to 660°C
TQSoft Automatic Calibration YES YES YES YES
Heating Time (Approximately) 23°C to 140°C 32 Minutes
-45°C to 140°C 45 Minutes
23°C to 155°C 44 Minutes
-30°C to 140°C 56 Minutes
35°C to 425°C 27 Minutes 50°C to 700°C 46 Minutes
Well Sizes Varied Varied Varied Varied
Physical Dimensions L x W x H 336 x 203 x 323mm 336 x 203 x 323mm 336 x 203 x 323mm 336 x 203 x 323mm
Certificate Provided Traceable Traceable Traceable Traceable
Carry Case Provided Sturdy, Wheeled Carry Case Sturdy, Wheeled Carry Case Sturdy, Wheeled Carry Case Sturdy, Wheeled Carry Case